Ethylene oxide is coming under increasing regulatory pressure from REACH, the planned PFAS ban, and the new Packaging Ordinance. As manufacturers reassess their materials, packaging and processes, they are encountering unresolved regulatory issues. TÜV SÜD is supporting the transition in its capacity as a Notified Body.
In the USA alone, around 20 billion medical devices undergo ethylene oxide (EO) sterilisation[1] each year. According to the FDA, this colourless gas is used in around half of all medical devices in the USA that must be sterilised. It can penetrate narrow cavities, complex geometries and multi-layered packaging. Yet it is still the sole option for many heat- or radiation-sensitive materials. However, the standard is losing ground – not in technical, but in regulatory terms.
A substance subject to various regulations
Under the CLP regulation, EO is classified as carcinogenic, mutagenic and toxic to reproduction (CMR). CMR substances of this kind are subject to particularly scrutiny under the EU’s chemicals regulation REACH. However, no complete substitution is possible yet. Although EO is classified as a substance of very high concern, its use in the context of medical devices will remain relevant in practical terms for the time being if no validated and regulatory-compliant alternatives are widely available.
Yet a growing need for adaptation is emerging from two further directions. The planned ban on PFAS identifies areas including medical device packaging as affected applications. Although the five national authorities, among them Germany, are providing for long transition periods [2], the trajectory is clear. At the same time the EU’s Packaging Regulation (PPWR) is establishing a new legal framework, compelling manufacturers to review their sterile barrier systems and reassess their packaging concepts at an early stage[3].
In the US, the Environmental Protection Agency (EPA) is tightening emission restrictions for sterilisation facilities. Its 2024 regulation increased the pressure on operators and manufacturers to adapt. However, the US government granted additional facility-specific exemptions in 2025,[4] and the EPA even published a formal “Proposed Reconsideration” in 2026. Debate is fuelled further by claims for damages and the cautious approach shown by insurers. However, the conclusion remains the same: the scope for ethylene oxide is narrowing.
Four alternative processes are gaining ground
In the meantime, various methods are maturing after undergoing further technical advancement. Vaporised hydrogen peroxide (VH₂O₂) usually operates under a vacuum; however, it is problematic with cellulose. Chlorine dioxide (ClO₂) also requires a vacuum and leaves potential residues of chlorites and chlorides. Nitrogen dioxide (NO₂) may require multiple pressure cycles depending on the application and can lead to discoloration of certain polymers. And low-energy electron beam sterilisation does not require gas and operates at normal pressure but may cause brittleness in some plastics.
Each of these methods offers advantages such as shorter cycle times or lower residue levels. Yet they may often be associated with a lack of reliable data on material compatibility, permeation and long-term stability. However, EO has been researched for decades and is covered by standards such as EN ISO 11135.
Given this, specific questions must be addressed before any changeover. Can the product withstand a deep vacuum? Are all product and packaging materials compatible with the sterilising agent? Is the throughput sufficient for the quantities planned? Can the supply chain keep pace with the changeover? Some early data is collected in scientific cross industrial efforts or joint efforts with FDA. Still – data in a comparative overview is not readily available.
Considerations on materials and packaging interactions
Every new method brings renewed focus to the interaction between process, material and packaging. Other sterilising agents attack many materials more harshly than EO, modifying the polymer structure, generating new degradation products or changing the profile of extractables and leachables. Effects of these kinds impact on biocompatibility, particle release and stability, including during storage and transport.
A further issue is the material substitution required by regulations. If a material containing PFAS is no longer used, the conditions for sterilisation change fundamentally. A material that was previously suitable for EO may behave differently following substitution. As a result, scrutiny once again turns to the validity of established processes. Sterilisation must therefore be regarded as part of an overall system spanning materials, manufacture, packaging and application.
A further key factor is penetration; some sterilising agents need to penetrate deep into multi-layered packaging to kill microorganisms reliably – a requirement that cannot be met by every design.
Need for a regulatory review
In regulatory terms, a change to the sterilisation process is often regarded as a significant change and may, under certain circumstances, necessitate a new MDR conformity assessment in line with MDCG 2020-3. Although the guidance document permits individual manufacturing steps to be transferred between production lines, this is only permissible where protocols have been validated and design outcomes remain unchanged. This extensive body of experience is as yet lacking for new sterilisation methods. Manufacturers and Notified Bodies therefore frequently assesses on a case-by-case basis when assessing what evidence is sufficient.
The framework is made up of several standards. EN ISO 14937 covers alternative methods in general. EN ISO 10993-1 governs biological evaluation, while EN ISO 10993-7 addresses residues from EO treatment as well as 10993-17 more generally addresses toxicological impact by residues.
Early decisions set the course for market readiness
Against this backdrop, many companies are making strategic decisions. While some are continuing to optimise their EO processes, others are already weighing alternatives. The issue is not merely one of technical feasibility but one of planning certainty in the coming years. Investments in equipment, adapting supply chains, changing packaging and validation need to pay off over the long term.
Consideration of design, materials and regulatory implications must take place at an early stage. Equally crucial is close coordination with sterilisation service providers, testing laboratories and Notified Bodies. The Industry Scientific Exchange (ISE) provides a platform for this. In Munich, ISE brought together over 75 experts from more than 30 organisations, with discussions that resulted in specific calls for action addressed to the industry. Uniform criteria for the functional equivalence of packaging materials are needed, as are peer-reviewed studies on material stability and standardised worst-case parameters for each process.
No full-scale replacement for EO appears to be likely soon. Instead, it is more likely that several methods will coexist, depending on the products and materials in question. For manufacturers, this will mean restructuring their development processes and integrating planning of sterilisation, material selection and packaging at an early stage. By setting the course of their operations now, they will secure some scope for action before EO comes under any further pressure.
Author
Dr. Jan Havel
Director Product Review Biological Safety of Medical Devices
Medical and Health Services
TÜV SÜD Product Service GmbH
Contact
TÜV SÜD Product Service
Ridlerstrasse 65
80339 Munich, Germany
jan.havel@tuevsued.de
[1] Figure of “around 20 billion medical devices per year”: AdvaMed, “Ethylene Oxide & Sterilisation”, advamed.org (based on FDA data).
[2] PFAS restriction proposal, submitted by Germany, the Netherlands, Sweden, Denmark and Norway; long exemptions, some time-limited, including for medical device packaging (e.g. 13.5 years for fluoropolymer packaging). Scientific assessment by the ECHA committees scheduled for 2026. Source: ECHA; Covington & Burling LLP, cov.com (2026)
[3] Packaging and Packaging Waste Regulation (PPWR), Regulation (EU) 2025/40; in force since 11 February 2025, applicable from 12 August 2026. Source: European Commission, environment.ec.europa.eu
[4] EPA emissions standard for sterilisation facilities dated April 2024; transition period extended by the US government in 2025. Source: The White House (Presidential Action, July 2025); FDA.