This recall involves removing certain devices from where they are used or sold. This recall also involves updated use instructions for devices not subject to removal. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without following the recommendations below.
Affected Product
The FDA is aware that Tandem Diabetes has issued a letter to affected customers recommending certain insulin pumps identified for replacement be removed from where they are used or sold. Tandem Diabetes also issued a separate letter providing updated use instructions for all other Tandem Mobi Insulin Pumps with Interoperable Technology.
Affected devices:
What to Do
Be aware that devices included in this recall have different recommendations based on the pump serial number. If your pump’s serial number is included in the “List of Affected Product Subject to Replacement,” follow recommendation number 1 below. If your pump’s serial number is not included in the “List of Affected Product Subject to Replacement,” follow recommendation number 2 below.
On September 23, 2026, Tandem Diabetes sent letters to customers recommending the following actions based on their pump’s serial number:
- Letter to customers whose pumps require replacement
- Confirm your pump is affected in the serial number look-up list above.
- Immediately submit a request for return and replacement of your pump using the link provided in the customer letter from Tandem Diabetes or by contacting Tandem Customer Technical Support at 877-801-6901.
- Always respond to an occlusion alarm and follow the recommended troubleshooting steps in the User Guide. Acknowledge the alarm, inspect the cartridge, tubing, and infusion site for signs of damage or blockage, correct any identified condition, and resume insulin delivery.
- If you continue to use the pump, also ensure that your Tandem Mobi mobile app has push notifications enabled so that if the alarm does occur, you will also be notified by the app. The pump will notify you of an occlusion alarm using audible tones or vibrations depending on your pump settings, progressing to louder audible tones until the alarm is acknowledged.
- Note: If you need assistance with enabling smartphone notifications to receive alerts from your mobile app, please access the article on this topic from Tandem’s support page.
- Regularly check your blood sugar to ensure you are not having unexpectedly high or low readings.
- Be prepared to use a backup method of insulin delivery as directed by your healthcare provider and described in your pump training and User Guide.
- Letter to all other Tandem Mobi users
- You may continue to use your pump while following the instructions below. Always respond to an occlusion alarm and follow the recommended troubleshooting steps in the User Guide. Acknowledge the alarm, inspect the cartridge, tubing, and infusion site for signs of damage or blockage, correct any identified condition, and resume insulin delivery.
- If you continue to use the pump, also ensure that your Tandem Mobi mobile app has push notifications enabled so that if the alarm does occur, you will also be notified by the app. The pump will notify you of an occlusion alarm using audible tones or vibrations depending on your pump settings, progressing to louder audible tones until the alarm is acknowledged.
- Note: If you need assistance with enabling smartphone notifications to receive alerts from your mobile app, please access the article on this topic from Tandem’s support page.
- If you start to experience an unusually high number of occlusion alarms, or occlusion alarms unexpectedly and repeatedly disrupt your therapy even after following troubleshooting steps, contact Tandem Customer Support.
- Be prepared to use a backup method of insulin delivery as directed by your healthcare provider and described in your pump training and User Guide.
Healthcare providers and United States distributors and pharmacies also received letters informing them of the recall to better serve their patients and customers.
Reason for Recall
Tandem Diabetes stated that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
As of October 28, 2026, Tandem Diabetes has reported 1,248 serious injuries and no deaths associated with this issue.
Device Use
The Tandem Mobi insulin pump is intended for subcutaneous delivery of insulin for diabetes management. The pump communicates with compatible devices and can be used with interoperable continuous glucose monitors and automated insulin dosing systems to automatically adjust insulin delivery.
Contact Information
Customers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact Tandem Diabetes at techsupport@tandemdiabetes.com or 877-801-6901.
Additional FDA Resources
Additional Company Resources
Company-provided information is posted here by the FDA as a public service.
Unique Device Identifier (UDI)
The unique device identifier (UDI) helps identify individual medical devices sold in the United States from distribution to use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified more quickly, and as a result, problems potentially resolved more quickly.
How do I report a problem?
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.